Choosing and using measuring apparatus

Edexcel IAL Chemistry · WCH13 Practical Skills in Chemistry I

Many marks in Unit 3 come from choosing the right piece of apparatus and using it properly. The choice depends on how precisely the quantity has to be known.

Practise this in the app: instant marking and a fix for every mistake. Free, no email needed to start.

Practise this lesson

Which apparatus for which job

ApparatusUseTypical resolution or uncertainty
volumetric (bulb) pipette, 25.0 cm3an accurately known fixed volume of solutionabout ±0.06 cm3
burette, 50 cm3variable volumes added accurately (titres, or a volume such as 7.50 cm3)read to 0.05 cm3; ±0.05 cm3 per reading
volumetric flask, 250.0 cm3making a standard solution of accurately known concentrationthe mark gives one exact volume
measuring cylinderapproximate volumes, such as an acid that is in excessabout ±0.5 cm3 for a 100 cm3 cylinder
gas syringe, 100 cm3volume of gas producedusually read to 1 cm3
balancemass2 or 3 decimal places (0.01 g or 0.001 g)
thermometertemperature0.1 °C (digital) to 1 °C (some glass ones)

Values follow Pearson's practical guide; a question will give the uncertainty if it matters.

Range and resolution

Using them properly

Worked example

Choose apparatus for each: (a) 25.0 cm3 of sodium hydroxide for a titration; (b) 20 cm3 of dilute sulfuric acid in excess; (c) about 80 cm3 of hydrogen gas; (d) 12.40 cm3 of potassium iodide solution.

  1. (a) A fixed, accurately known volume: 25.0 cm3 volumetric pipette.
  2. (b) In excess, so precision is not needed: measuring cylinder.
  3. (c) Gas of about 80 cm3: 100 cm3 gas syringe.
  4. (d) A variable volume known to 0.05 cm3: burette.

Answer: pipette; measuring cylinder; 100 cm3 gas syringe; burette

Key idea: Pipette for fixed accurate volumes, burette for variable accurate volumes, volumetric flask for standard solutions, measuring cylinder only where precision does not matter. Check range and resolution, and weigh by difference.

Check yourself

  1. Which apparatus should be used to add exactly 7.50 cm3 of a solution to a reaction mixture?
    • A 10 cm3 measuring cylinder
    • A 25.0 cm3 pipette
    • A beaker with volume markings
    • A burette
    Show answer

    A burette. A burette delivers variable volumes to 0.05 cm3. Pipettes give only their fixed volume.

  2. Dilute sulfuric acid is added in excess before a titration. Why is a measuring cylinder acceptable for it?
    • Sulfuric acid damages pipettes
    • Its exact amount does not affect the result, because it is in excess
    • Measuring cylinders are more accurate than pipettes
    • The acid is the solution being titrated
    Show answer

    Its exact amount does not affect the result, because it is in excess. Only the limiting quantities must be known precisely.

  3. About 85 cm3 of gas is expected. Which piece of apparatus is best?
    • A 50 cm3 gas syringe
    • A 10 cm3 measuring cylinder
    • A 25.0 cm3 pipette
    • A 100 cm3 gas syringe
    Show answer

    A 100 cm3 gas syringe. The range must cover the expected volume.

  4. What should a pipette be rinsed with just before it is used?
    • The solution it is going to deliver
    • Distilled water
    • Ethanol
    • The solution in the burette
    Show answer

    The solution it is going to deliver. Water left inside would dilute the solution and lower the moles delivered.

  5. A balance shows 12.37 g. What is its resolution, in g?
    Show answer

    0.01 g. It reads to two decimal places, so the smallest change it shows is 0.01 g.

  6. Why is a solid 'weighed by difference' when making a standard solution?
    • The mass found is what actually left the container, even if some solid stays behind
    • It makes the balance more precise
    • It removes the need to use a volumetric flask
    • It stops the solid absorbing water
    Show answer

    The mass found is what actually left the container, even if some solid stays behind. Reweighing the container accounts for any solid that sticks to it.

  7. Explain why the funnel should be removed from the top of a burette before the initial reading is taken.
    Show answer

    Drops of solution left in the funnel could fall into the burette and change the reading.. Liquid draining from the funnel would make the initial reading wrong.