Chromatography and high-resolution mass spectrometry
Chromatography separates the components of a mixture, and high-resolution mass spectrometry identifies them by their accurate relative molecular mass. Together they are used in forensic science and drug testing.
How chromatography works (spec 15.21)
Every type has a stationary phase (which does not move) and a mobile phase (which moves through it). Components separate because they have different relative affinities: one that is more strongly attracted to (or adsorbed on) the stationary phase, or less soluble in the mobile phase, moves more slowly.
Paper and thin-layer chromatography (spec 15.22)
In TLC the stationary phase is a thin layer of silica or alumina on a plate; in paper chromatography it is the paper (with water held in it). The solvent is the mobile phase.
Rf = distance moved by the spot / distance moved by the solvent front
Rf is always less than 1 and, for given conditions, is characteristic of a compound. Colourless spots are shown with UV light or a locating agent.
HPLC and GC (spec 15.23)
High-performance liquid chromatography pumps a liquid mobile phase at high pressure through a column packed with a solid. Gas chromatography carries a vaporised sample in an inert gas (such as helium or nitrogen) through a column whose stationary phase is a liquid on a solid support. Each component is identified by its retention time: the time from injection to detection. Peak areas show relative amounts. Coupled to a mass spectrometer (GC-MS or HPLC-MS), each separated component is identified from its mass spectrum: used in forensics, environmental analysis and testing athletes for banned drugs.
Accurate Mr (spec 15.17)
A high-resolution mass spectrometer measures m/z to four decimal places. Using accurate isotopic masses (1H 1.0078, 12C 12.0000, 14N 14.0031, 16O 15.9949), molecules with the same integer Mr can be told apart: CO 27.9949, N2 28.0062 and C2H4 28.0312.
Worked example
A molecular ion has m/z 60.0573. Is the compound C3H8O or C2H4O2?
- C3H8O: 3(12.0000) + 8(1.0078) + 15.9949 = 60.0573
- C2H4O2: 2(12.0000) + 4(1.0078) + 2(15.9949) = 60.0210
Answer: C3H8O
Check yourself
- A spot moves 3.2 cm and the solvent front 8.0 cm. Calculate Rf.
Show answer
0.40. Rf = 3.2 / 8.0 = 0.40.
- In thin-layer chromatography, what is the stationary phase?
- The solvent
- A thin layer of silica or alumina on a plate
- The sample
- The air around the plate
Show answer
A thin layer of silica or alumina on a plate. The solid coating stays still while the solvent moves up the plate.
- Why do two compounds have different Rf values on the same plate?
- They have different colours
- They were spotted at different heights
- One is heavier
- They have different relative affinities for the stationary and mobile phases
Show answer
They have different relative affinities for the stationary and mobile phases. A compound adsorbed more strongly or less soluble in the solvent moves less.
- In gas chromatography, what is used to identify a component?
- Its retention time
- Its Rf value
- Its colour
- Its peak height only
Show answer
Its retention time. Retention time is characteristic under fixed conditions.
- Calculate the accurate relative molecular mass of C3H8O to four decimal places. (1H 1.0078, 12C 12.0000, 16O 15.9949)
Show answer
60.0573. 3(12.0000) + 8(1.0078) + 15.9949 = 60.0573.
- A molecular ion peak is at m/z 28.0312. Which molecule is it? (1H 1.0078, 12C 12.0000, 14N 14.0031, 16O 15.9949)
- C2H4
- CO
- N2
- CH2N
Show answer
C2H4. C2H4 = 24.0000 + 4.0312 = 28.0312; CO = 27.9949; N2 = 28.0062.
- Why is gas chromatography often coupled with mass spectrometry in drug testing?
- MS separates the mixture and GC identifies each part
- GC measures the accurate Mr directly
- GC separates the mixture and MS identifies each separated component
- It avoids needing a mobile phase
Show answer
GC separates the mixture and MS identifies each separated component. Retention times alone may be shared by different compounds; the mass spectrum confirms identity.